ANALYTICAL METHOD VALIDATION
By Ch.Sandhya
Validation is defined as the process of establishing a
documented evidence that a specific process will consistently
produce a product meeting its predetermined specifications
and quality attributes.
The analytical procedure refers to the way of performing
the analysis. It should describe in detail the steps necessary to
perform each analytical test. This may include but is not
limited to the sample, the reference standard and the reagents
preparations, use of the apparatus, generation of the
calibration curve, use of the formulae for the calculation etc.
1. SPECIFICITY
SPECIFICITY is the ability to assess unequivocally the analyte in
presence of components which may be expected to be present.
An investigation of specificity should be conducted during the validation of
identification tests, the determination of impurities and the assay. The
procedures used to demonstrate specificity will depend on the intended
objective of the analytical procedure.
1.1. Identification
Suitable identification tests should be able to discriminate between
compounds of closely related structures which are likely to be present. The
discrimination of a procedure may be confirmed by obtaining positive
results (perhaps by comparison with a known reference material) from
samples containing the analyte, coupled with negative results from samples
which do not contain the analyte
In addition, the identification test may be applied to materials structurally
similar to or closely related to the analyte to confirm that a positive
response is not obtained. The choice of such potentially interfering
materials should be based on sound scientific judgement with a
consideration of the interferences that could occur.
1.2. Assay and Impurity Test(s)
For chromatographic procedures, representative chromatograms should be
used to demonstrate specificity and individual components should be
appropriately labelled. Similar considerations should be given to other
separation techniques.
Critical separations in chromatography should be investigated at an
appropriate level. For critical separations, specificity can be demonstrated
by the resolution of the two components which elute closest to each other.
In cases where a non-specific assay is used, other supporting analytical
procedures should be used to demonstrate overall specificity. For example,
where a titration is adopted to assay the drug substance for release, the
combination of the assay and a suitable test for impurities can be used.
The approach is similar for both assay and impurity tests:
1.2.1 Impurities are available
For the assay , this should involve demonstration of the discrimination of
the analyte in the presence of impurities and/or excipients; practically, this
can be done by spiking pure substances (drug substance or drug product)
with appropriate levels of impurities and/or excipients and demonstrating
that the assay result is unaffected by the presence of these materials (by
comparison with the assay result obtained on unspiked samples).
1.2.2 Impurities are not available
If impurity or degradation product standards are unavailable, specificity
may be demonstrated by comparing the test results of samples containing
impurities or degradation products to a second well-characterized
procedure e.g.: pharmacopoeial method or other validated analytical
procedure (independent procedure). As appropriate, this should include
samples stored under relevant stress conditions: light, heat, humidity,
acid/base hydrolysis and oxidation.
-for the assay, the two results should be compared;
- for the impurity tests, the impurity profiles should be compared.
- Peak purity tests may be useful to show that the analyte chromatographic
peak is not attributable to more than one component (e.g., diode array,
mass spectrometry).
2. LINEARITY
LINEARITY of an analytical procedure is its ability (within a given range)
to obtain test results which are directly proportional to the concentration
(amount) of analyte in the sample.
A linear relationship should be evaluated across the range (see section 3) of
the analytical procedure. It may be demonstrated directly on the drug
substance (by dilution of a standard stock solution) and/or separate
weighings of synthetic mixtures of the drug product components, using the
proposed procedure. The latter aspect can be studied during investigation
of the range.
Linearity should be evaluated by visual inspection of a plot of signals as a
function of analyte concentration or content.
If there is a linear relationship, test results should be
evaluated by appropriate statistical methods, for example, by
calculation of a regression line by the method of least squares.
The correlation coefficient, y-intercept, slope of the
regression line and residual sum of squares should be
submitted. A plot of the data should be included. In addition,
an analysis of the deviation of the actual data points from the
regression line may also be helpful for evaluating linearity.
For the establishment of linearity, a minimum of 5
concentrations is recommended. Other approaches should be
justified.
3. RANGE
RANGE of an analytical procedure is the interval between the upper and
lower concentration (amounts) of analyte in the sample (including these
concentrations) for which it has been demonstrated that the analytical
procedure has a suitable level of precision, accuracy and linearity
The specified range is normally derived from linearity studies and depends
on the intended application of the procedure. It is established by
confirming that the analytical procedure provides an acceptable degree of
linearity, accuracy and precision when applied to samples containing
amounts of analyte within or at the extremes of the specified range of the
analytical procedure.
The following minimum specified ranges should be considered:
- for the assay of a drug substance or a finished (drug) product: normally
from 80 to 120 percent of the test concentration;
- for content uniformity, covering a minimum of 70 to 130 percent of the
test concentration, unless a wider more appropriate range, based on the
nature of the dosage form (e.g., metered dose inhalers), is justified;
- for dissolution testing: +/-20 % over the specified range;
4. ACCURACY
ACCURACY of an analytical method is the closeness of test results
obtained by that method to the true value.
Accuracy should be established across the specified range of the analytical
procedure.
4.1. Assay
4.1.1 Drug Substance
Several methods of determining accuracy are available:
a) application of an analytical procedure to an analyte of known purity
b) comparison of the results of the proposed analytical procedure with
those of a second well-characterized procedure, the accuracy of which is
stated and/or defined
c) accuracy may be inferred once precision, linearity and specificity have
been established.
4.1.2 Drug Product
Several methods for determining accuracy are available:
a) application of the analytical procedure to synthetic mixtures of the drug
product components to which known quantities of the drug substance to be
analysed have been added;
b) in cases where it is impossible to obtain samples of all drug product
components , it may be acceptable either to add known quantities of the
analyte to the drug product or to compare the results obtained from a
second, well characterized procedure, the accuracy of which is stated
and/or defined
c) accuracy may be inferred once precision, linearity and specificity have
been established.
4.2. Impurities (Quantitation)
Accuracy should be assessed on samples (drug substance/drug product)
spiked with known amounts of impurities.
In cases where it is impossible to obtain samples of certain impurities
and/or degradation products, it is considered acceptable to compare results
obtained by an independent procedure. The response factor of the drug
substance can be used.
It should be clear how the individual or total impurities are to be
determined e.g., weight/weight or area percent, in all cases with respect to
the major analyte.
4.3. Recommended Data
Accuracy should be assessed using a minimum of 9 determinations over a
minimum of 3 concentration levels covering the specified range (e.g., 3
concentrations/3 replicates each of the total analytical procedure).
Accuracy should be reported as percent recovery by the assay of known
added amount of analyte in the sample or as the difference between the
mean and the accepted true value together with the confidence intervals.
5. PRECISION
PRECISION of an analytical method is the degree of agreement among
individual test results when the method is applied repeatedly to multiple
samplings of a homogenous sample.
Validation of tests for assay and for quantitative determination of
impurities includes an investigation of precision.
5.1. Repeatability It refers to the results of the method operating over a
short time interval under the same conditions.
Repeatability should be assessed using:
a) a minimum of 9 determinations covering the specified range for the
procedure (e.g., 3 concentrations/3 replicates each); or
b) a minimum of 6 determinations at 100% of the test concentration.
5.2. Intermediate Precision
It refers to the results from within lab variations due to random events
such as differences in experimental periods, analysts, equipment.
The extent to which intermediate precision should be established
depends on the circumstances under which the procedure is intended to be
used. The applicant should establish the effects of random events on the
precision of the analytical procedure.
Typical variations to be studied include days, analysts, equipment, etc. It
is not considered necessary to study these effects individually. The use of an
experimental design (matrix) is encouraged.
5.3. Reproducibility
It refers to the results of collaborative studies among laboratories.
Reproducibility is assessed by means of an inter-laboratory trial.
Reproducibility should be considered in case of the standardization of an
analytical procedure, for instance, for inclusion of procedures in
pharmacopoeias. These data are not part of the marketing authorization
dossier.
5.4. Recommended Data
The standard deviation, relative standard deviation (coefficient of
variation) and confidence interval should be reported for each type of
precision investigated.
6. DETECTION LIMIT
DETECTION LIMIT of an individual analytical procedure is the lowest
amount of analyte in a sample which can be detected but not necessarily
quantitated, under the stated experimental conditions.
Several approaches for determining the detection limit are possible,
depending on whether the procedure is a non-instrumental or
instrumental. Approaches other than those listed below may be acceptable.
6.1. Based on Visual Evaluation
Visual evaluation may be used for non-instrumental methods but may also
be used with instrumental methods.
The detection limit is determined by the analysis of samples with known
concentrations of analyte and by establishing the minimum level at which
the analyte can be reliably detected.
6.2. Based on Signal-to-Noise
This approach can only be applied to analytical procedures which exhibit
baseline noise.
Determination of the signal-to-noise ratio is performed by comparing
measured signals from samples with known low concentrations of analyte
with those of blank samples and establishing the minimum concentration
at which the analyte can be reliably detected. A signal-to-noise ratio
between 3 or 2:1 is generally considered acceptable for estimating the
detection limit.
6.3 Based on the Standard Deviation of the Response and the
Slope
The detection limit (DL) may be expressed as: DL = 3.3 σ/ S
where σ = the standard deviation of the response
S = the slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte. The
estimate of σ may be carried out in a variety of ways, for example:
6.3.1 Based on the Standard Deviation of the Blank
Measurement of the magnitude of analytical background response is
performed by analyzing an appropriate number of blank samples and
calculating the standard deviation of these responses.
6.3.2 Based on the Calibration Curve
A specific calibration curve should be studied using samples containing an
analyte in the range of DL. The residual standard deviation of a regression
line or the standard deviation of y-intercepts of regression lines may be
used as the standard deviation.
6.4 Recommended Data
The detection limit and the method used for determining the detection limit
should be presented. If DL is determined based on visual evaluation or
based on signal to noise ratio, the presentation of the relevant
chromatograms is considered acceptable for justification.
7. QUANTITATION LIMIT
QUANTITATION LIMIT of an individual analytical procedure is the
lowest amount of analyte in a sample which can be quantitatively
determined with suitable precision and accuracy
Several approaches for determining the quantitation limit are possible,
depending on whether the procedure is a non-instrumental or
instrumental. Approaches other than those listed below may be acceptable.
7.1. Based on Visual Evaluation
Visual evaluation may be used for non-instrumental methods but may also
be used with instrumental methods.
The quantitation limit is generally determined by the analysis of samples
with known concentrations of analyte and by establishing the minimum
level at which the analyte can be quantified with acceptable accuracy and
precision.
7.2. Based on Signal-to-Noise Approach
This approach can only be applied to analytical procedures that exhibit
baseline noise.
Determination of the signal-to-noise ratio is performed by comparing
measured signals from samples with known low concentrations of analyte
with those of blank samples and by establishing the minimum
concentration at which the analyte can be reliably quantified. A typical
signal-to-noise ratio is 10:1.
7.3. Based on the Standard Deviation of the Response and the
Slope
The quantitation limit (QL) may be expressed as: QL = 10 σ /S
where σ = the standard deviation of the response
S = the slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte.
The estimate of σ may be carried out in a variety of ways for example:
7.3.1 Based on Standard Deviation of the Blank
Measurement of the magnitude of analytical background response is
performed by analyzing an appropriate number of blank samples and
calculating the standard deviation of these responses.
7.3.2 Based on the Calibration Curve
A specific calibration curve should be studied using samples, containing an
analyte in the range of QL. The residual standard deviation of a regression
line or the standard deviation of y-intercepts of regression lines may be
used as the standard deviation.
7.4 Recommended Data
The QL and the method used for determining the QL should be presented.
The limit should be subsequently validated by the analysis of a suitable
number of samples known to be near or prepared at the quantitation limit.
8. ROBUSTNESS
ROBUSTNESS of an analytical procedure is a measure of its capacity to
remain unaffected by small, but deliberate variations in method parameters
and provides an indication of its reliability during normal usage.
The evaluation of robustness should be considered during the development
phase and depends on the type of procedure under study. It should show
the reliability of an analysis with respect to deliberate variations in method
parameters.
If measurements are susceptible to variations in analytical conditions, the
analytical conditions should be suitably controlled or a precautionary
statement should be included in the procedure. One consequence of the
evaluation of robustness should be that a series of system suitability
parameters (e.g., resolution test) is established to ensure that the validity of
the analytical procedure is maintained whenever used.
Examples of typical variations are:
- stability of analytical solutions;
- extraction time.
In the case of liquid chromatography, examples of typical variations are:
- influence of variations of pH in a mobile phase;
- influence of variations in mobile phase composition;
- different columns (different lots and/or suppliers);
- temperature;
- flow rate.
In the case of gas-chromatography, examples of typical variations are:
- different columns (different lots and/or suppliers);
- temperature;
- flow rate.
9. SYSTEM SUITABILITY TESTING
System suitability testing is an integral part of many analytical procedures.
The tests are based on the concept that the equipment, electronics,
analytical operations and samples to be analyzed constitute an integral
system that can be evaluated as such. System suitability test parameters to
be established for a particular procedure depend on the type of procedure
being validated. See Pharmacopoeias for additional information.
RUGGEDNESS of an analytical method is the degree of reproducibility of
test results obtained by the analysis of the same samples under a variety of
conditions, such as different laboratories different analyst, different
instruments, different lots of reagent, different elapsed assay times,
different assay temperatures, different days, etc.
It is included in USP but not included in ICH

Validation parameters

  • 1.
  • 2.
    Validation is definedas the process of establishing a documented evidence that a specific process will consistently produce a product meeting its predetermined specifications and quality attributes. The analytical procedure refers to the way of performing the analysis. It should describe in detail the steps necessary to perform each analytical test. This may include but is not limited to the sample, the reference standard and the reagents preparations, use of the apparatus, generation of the calibration curve, use of the formulae for the calculation etc.
  • 3.
    1. SPECIFICITY SPECIFICITY isthe ability to assess unequivocally the analyte in presence of components which may be expected to be present. An investigation of specificity should be conducted during the validation of identification tests, the determination of impurities and the assay. The procedures used to demonstrate specificity will depend on the intended objective of the analytical procedure. 1.1. Identification Suitable identification tests should be able to discriminate between compounds of closely related structures which are likely to be present. The discrimination of a procedure may be confirmed by obtaining positive results (perhaps by comparison with a known reference material) from samples containing the analyte, coupled with negative results from samples which do not contain the analyte
  • 4.
    In addition, theidentification test may be applied to materials structurally similar to or closely related to the analyte to confirm that a positive response is not obtained. The choice of such potentially interfering materials should be based on sound scientific judgement with a consideration of the interferences that could occur. 1.2. Assay and Impurity Test(s) For chromatographic procedures, representative chromatograms should be used to demonstrate specificity and individual components should be appropriately labelled. Similar considerations should be given to other separation techniques. Critical separations in chromatography should be investigated at an appropriate level. For critical separations, specificity can be demonstrated by the resolution of the two components which elute closest to each other.
  • 5.
    In cases wherea non-specific assay is used, other supporting analytical procedures should be used to demonstrate overall specificity. For example, where a titration is adopted to assay the drug substance for release, the combination of the assay and a suitable test for impurities can be used. The approach is similar for both assay and impurity tests: 1.2.1 Impurities are available For the assay , this should involve demonstration of the discrimination of the analyte in the presence of impurities and/or excipients; practically, this can be done by spiking pure substances (drug substance or drug product) with appropriate levels of impurities and/or excipients and demonstrating that the assay result is unaffected by the presence of these materials (by comparison with the assay result obtained on unspiked samples).
  • 6.
    1.2.2 Impurities arenot available If impurity or degradation product standards are unavailable, specificity may be demonstrated by comparing the test results of samples containing impurities or degradation products to a second well-characterized procedure e.g.: pharmacopoeial method or other validated analytical procedure (independent procedure). As appropriate, this should include samples stored under relevant stress conditions: light, heat, humidity, acid/base hydrolysis and oxidation. -for the assay, the two results should be compared; - for the impurity tests, the impurity profiles should be compared. - Peak purity tests may be useful to show that the analyte chromatographic peak is not attributable to more than one component (e.g., diode array, mass spectrometry).
  • 7.
    2. LINEARITY LINEARITY ofan analytical procedure is its ability (within a given range) to obtain test results which are directly proportional to the concentration (amount) of analyte in the sample. A linear relationship should be evaluated across the range (see section 3) of the analytical procedure. It may be demonstrated directly on the drug substance (by dilution of a standard stock solution) and/or separate weighings of synthetic mixtures of the drug product components, using the proposed procedure. The latter aspect can be studied during investigation of the range. Linearity should be evaluated by visual inspection of a plot of signals as a function of analyte concentration or content.
  • 8.
    If there isa linear relationship, test results should be evaluated by appropriate statistical methods, for example, by calculation of a regression line by the method of least squares. The correlation coefficient, y-intercept, slope of the regression line and residual sum of squares should be submitted. A plot of the data should be included. In addition, an analysis of the deviation of the actual data points from the regression line may also be helpful for evaluating linearity. For the establishment of linearity, a minimum of 5 concentrations is recommended. Other approaches should be justified.
  • 9.
    3. RANGE RANGE ofan analytical procedure is the interval between the upper and lower concentration (amounts) of analyte in the sample (including these concentrations) for which it has been demonstrated that the analytical procedure has a suitable level of precision, accuracy and linearity The specified range is normally derived from linearity studies and depends on the intended application of the procedure. It is established by confirming that the analytical procedure provides an acceptable degree of linearity, accuracy and precision when applied to samples containing amounts of analyte within or at the extremes of the specified range of the analytical procedure.
  • 10.
    The following minimumspecified ranges should be considered: - for the assay of a drug substance or a finished (drug) product: normally from 80 to 120 percent of the test concentration; - for content uniformity, covering a minimum of 70 to 130 percent of the test concentration, unless a wider more appropriate range, based on the nature of the dosage form (e.g., metered dose inhalers), is justified; - for dissolution testing: +/-20 % over the specified range;
  • 11.
    4. ACCURACY ACCURACY ofan analytical method is the closeness of test results obtained by that method to the true value. Accuracy should be established across the specified range of the analytical procedure. 4.1. Assay 4.1.1 Drug Substance Several methods of determining accuracy are available: a) application of an analytical procedure to an analyte of known purity b) comparison of the results of the proposed analytical procedure with those of a second well-characterized procedure, the accuracy of which is stated and/or defined c) accuracy may be inferred once precision, linearity and specificity have been established.
  • 12.
    4.1.2 Drug Product Severalmethods for determining accuracy are available: a) application of the analytical procedure to synthetic mixtures of the drug product components to which known quantities of the drug substance to be analysed have been added; b) in cases where it is impossible to obtain samples of all drug product components , it may be acceptable either to add known quantities of the analyte to the drug product or to compare the results obtained from a second, well characterized procedure, the accuracy of which is stated and/or defined c) accuracy may be inferred once precision, linearity and specificity have been established.
  • 13.
    4.2. Impurities (Quantitation) Accuracyshould be assessed on samples (drug substance/drug product) spiked with known amounts of impurities. In cases where it is impossible to obtain samples of certain impurities and/or degradation products, it is considered acceptable to compare results obtained by an independent procedure. The response factor of the drug substance can be used. It should be clear how the individual or total impurities are to be determined e.g., weight/weight or area percent, in all cases with respect to the major analyte.
  • 14.
    4.3. Recommended Data Accuracyshould be assessed using a minimum of 9 determinations over a minimum of 3 concentration levels covering the specified range (e.g., 3 concentrations/3 replicates each of the total analytical procedure). Accuracy should be reported as percent recovery by the assay of known added amount of analyte in the sample or as the difference between the mean and the accepted true value together with the confidence intervals.
  • 15.
    5. PRECISION PRECISION ofan analytical method is the degree of agreement among individual test results when the method is applied repeatedly to multiple samplings of a homogenous sample. Validation of tests for assay and for quantitative determination of impurities includes an investigation of precision. 5.1. Repeatability It refers to the results of the method operating over a short time interval under the same conditions. Repeatability should be assessed using: a) a minimum of 9 determinations covering the specified range for the procedure (e.g., 3 concentrations/3 replicates each); or b) a minimum of 6 determinations at 100% of the test concentration.
  • 16.
    5.2. Intermediate Precision Itrefers to the results from within lab variations due to random events such as differences in experimental periods, analysts, equipment. The extent to which intermediate precision should be established depends on the circumstances under which the procedure is intended to be used. The applicant should establish the effects of random events on the precision of the analytical procedure. Typical variations to be studied include days, analysts, equipment, etc. It is not considered necessary to study these effects individually. The use of an experimental design (matrix) is encouraged.
  • 17.
    5.3. Reproducibility It refersto the results of collaborative studies among laboratories. Reproducibility is assessed by means of an inter-laboratory trial. Reproducibility should be considered in case of the standardization of an analytical procedure, for instance, for inclusion of procedures in pharmacopoeias. These data are not part of the marketing authorization dossier. 5.4. Recommended Data The standard deviation, relative standard deviation (coefficient of variation) and confidence interval should be reported for each type of precision investigated.
  • 18.
    6. DETECTION LIMIT DETECTIONLIMIT of an individual analytical procedure is the lowest amount of analyte in a sample which can be detected but not necessarily quantitated, under the stated experimental conditions. Several approaches for determining the detection limit are possible, depending on whether the procedure is a non-instrumental or instrumental. Approaches other than those listed below may be acceptable. 6.1. Based on Visual Evaluation Visual evaluation may be used for non-instrumental methods but may also be used with instrumental methods. The detection limit is determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be reliably detected.
  • 19.
    6.2. Based onSignal-to-Noise This approach can only be applied to analytical procedures which exhibit baseline noise. Determination of the signal-to-noise ratio is performed by comparing measured signals from samples with known low concentrations of analyte with those of blank samples and establishing the minimum concentration at which the analyte can be reliably detected. A signal-to-noise ratio between 3 or 2:1 is generally considered acceptable for estimating the detection limit.
  • 20.
    6.3 Based onthe Standard Deviation of the Response and the Slope The detection limit (DL) may be expressed as: DL = 3.3 σ/ S where σ = the standard deviation of the response S = the slope of the calibration curve The slope S may be estimated from the calibration curve of the analyte. The estimate of σ may be carried out in a variety of ways, for example: 6.3.1 Based on the Standard Deviation of the Blank Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses.
  • 21.
    6.3.2 Based onthe Calibration Curve A specific calibration curve should be studied using samples containing an analyte in the range of DL. The residual standard deviation of a regression line or the standard deviation of y-intercepts of regression lines may be used as the standard deviation. 6.4 Recommended Data The detection limit and the method used for determining the detection limit should be presented. If DL is determined based on visual evaluation or based on signal to noise ratio, the presentation of the relevant chromatograms is considered acceptable for justification.
  • 22.
    7. QUANTITATION LIMIT QUANTITATIONLIMIT of an individual analytical procedure is the lowest amount of analyte in a sample which can be quantitatively determined with suitable precision and accuracy Several approaches for determining the quantitation limit are possible, depending on whether the procedure is a non-instrumental or instrumental. Approaches other than those listed below may be acceptable. 7.1. Based on Visual Evaluation Visual evaluation may be used for non-instrumental methods but may also be used with instrumental methods. The quantitation limit is generally determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be quantified with acceptable accuracy and precision.
  • 23.
    7.2. Based onSignal-to-Noise Approach This approach can only be applied to analytical procedures that exhibit baseline noise. Determination of the signal-to-noise ratio is performed by comparing measured signals from samples with known low concentrations of analyte with those of blank samples and by establishing the minimum concentration at which the analyte can be reliably quantified. A typical signal-to-noise ratio is 10:1. 7.3. Based on the Standard Deviation of the Response and the Slope The quantitation limit (QL) may be expressed as: QL = 10 σ /S where σ = the standard deviation of the response S = the slope of the calibration curve The slope S may be estimated from the calibration curve of the analyte.
  • 24.
    The estimate ofσ may be carried out in a variety of ways for example: 7.3.1 Based on Standard Deviation of the Blank Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses. 7.3.2 Based on the Calibration Curve A specific calibration curve should be studied using samples, containing an analyte in the range of QL. The residual standard deviation of a regression line or the standard deviation of y-intercepts of regression lines may be used as the standard deviation. 7.4 Recommended Data The QL and the method used for determining the QL should be presented. The limit should be subsequently validated by the analysis of a suitable number of samples known to be near or prepared at the quantitation limit.
  • 25.
    8. ROBUSTNESS ROBUSTNESS ofan analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage. The evaluation of robustness should be considered during the development phase and depends on the type of procedure under study. It should show the reliability of an analysis with respect to deliberate variations in method parameters. If measurements are susceptible to variations in analytical conditions, the analytical conditions should be suitably controlled or a precautionary statement should be included in the procedure. One consequence of the evaluation of robustness should be that a series of system suitability parameters (e.g., resolution test) is established to ensure that the validity of the analytical procedure is maintained whenever used.
  • 26.
    Examples of typicalvariations are: - stability of analytical solutions; - extraction time. In the case of liquid chromatography, examples of typical variations are: - influence of variations of pH in a mobile phase; - influence of variations in mobile phase composition; - different columns (different lots and/or suppliers); - temperature; - flow rate. In the case of gas-chromatography, examples of typical variations are: - different columns (different lots and/or suppliers); - temperature; - flow rate.
  • 27.
    9. SYSTEM SUITABILITYTESTING System suitability testing is an integral part of many analytical procedures. The tests are based on the concept that the equipment, electronics, analytical operations and samples to be analyzed constitute an integral system that can be evaluated as such. System suitability test parameters to be established for a particular procedure depend on the type of procedure being validated. See Pharmacopoeias for additional information.
  • 28.
    RUGGEDNESS of ananalytical method is the degree of reproducibility of test results obtained by the analysis of the same samples under a variety of conditions, such as different laboratories different analyst, different instruments, different lots of reagent, different elapsed assay times, different assay temperatures, different days, etc. It is included in USP but not included in ICH